Margin evaluation of translational and rotational set-up errors in intensity modulated radiotherapy for cervical cancer

Abstract

A clinical target volume (CTV) to planning target volume (PTV) margin recipes was routinely used to ensure dose was actually delivered to target for all (most) patients. Currently used margin recipes were associated with only translational set-up errors in radiotherapy. However, when set-up errors extended to six-degree (6D) scope (three translational and three rotational set-up errors), margin recipe should be re-evaluated. The purpose of this study was to investigate dosimetric changes of targets (both CTV and PTV) coverage when 6D set-up errors were introduced and testify the practicability of currently used margin recipe in radiotherapy. A total number of 105 cone beam computer tomography scans for ten patients with cervical cancer were derived prior to treatment delivery and 6D set-up errors were acquired with image registration tools. Target coverage was evaluated retrospectively for 6D set-up errors introduced plan with 6 mm CTV to PTV margin. Target coverage of PTV showed significant decreases (3.3 %) in set-up errors introduced plans compared with original plans. But CTV coverage was not susceptible to these set-up errors. A tendency of coverage decrease for PTV along with distance away from treatment was testified, from −0.2 to −6.2 %. However, CTV seems changed less, from −0.2 to −0.8 %. The result indicate that a CTV to PTV margin of 6 mm was sufficient to take into account 6D set-up errors for most patients with cervical cancer. Future research suggests a smaller margin to further improve both tumor coverage and organs at risk sparing.

Background

External beam radiotherapy (EBRT) followed by brachytherapy is the standard treatment for cervical cancer, especially for advanced disease. Intensity modulated radiotherapy (IMRT) is increasingly used for EBRT of cervical cancer for its satisfying dose distribution on a plan to ensure target dose and limit normal tissue irradiation (Chen et al. 2007; Brixey et al. 2002; van de Bunt et al. 2006; Ahamad et al. 2005). However, on the other hand, the more accurate dose distribution of IMRT, the more precise treatment procedure is needed. Positional uncertainties and organ motion are two major factors contribute to treatment accuracy in EBRT (van Herk 2004; Beard et al. 1996; Antolak et al. 1998). Safety margin from CTV to PTV is a standard approach to ensure that the delivery region is adequate despite uncertainties in treatment.

Image guided radiotherapy (IGRT) provides an excellent solution to quantify and correct for patient set-up errors in modern radiotherapy (Zeidan et al. 2007; Mackie et al. 2003; Letourneau et al. 2005; Verellen et al. 2003). The IGRT equipment mounts an x-ray tube and a flat panel detector, CBCT, on linear accelerator. Patients have daily CBCT scan prior to treatment delivery. By sort of image registration method, one can estimate patient pose difference in the treatment room from the planning CT. With this advanced imaging techniques, 6D set-up errors (3 translational and 3 rotational, pitch, roll and yaw) can be detected and derived. Unfortunately, rotational set-up errors are not routinely corrected because most couches currently available can only correct one degree of these rotational movements.

The impact of rotational set-up errors is an attractive issue for researchers (van Herk 2004; Guckenberger et al. 2006; Kim et al. 2008; Lim et al. 2009; Peng et al. 2013). Cervical cancer target has the characteristic that it always has large cranio-caudal extension. Therefore, the target coverage is more susceptible to rotational movements especially in pitch rotation. There are some studies reported that larger discrepancies generated by pitch errors were found in cervical cancers (Laursen et al. 2012; Quint et al. 2001; Kaiser et al. 2006). A CTV to PTV margin recipes (2.5 SD of systematic translational set-up errors plus 0.7 SD of random translational set-up errors) introduced by van Herk et al. is Commonly used regardless of rotational set-up errors and shape of target (Lips et al. 2009). Whether these discrepancies are significant in dosimeteric for cervical cancers, should be evaluated carefully.

Rotational movements can be compensated by the method of using gantry and collimator rotations to correct for target movements (Yue et al. 2006). It is an alternative method of applying 6D repositioning device. This is also an ideal method to reestablished accumulated dose of errors introduced plan in treatment planning system (TPS). The purpose of this study is to evaluate the impact of correct rotational set-up errors in dosimeteric with clinical obtained data from multi-fractions.

Methods

Patient selection and plan

Ten cervical cancer patents undergoing definitive radiotherapy in Zhejiang Cancer Hospital during January 2012 and June 2012 were retrospectively reviewed. Planning CT scans were performed with Philips CT-sim scanner with 5-mm slice thickness (Philips Medical System, Eindhoven, Netherlands). Patients were treated in supine position and placed on room laser system according to markers on body. All patients used customized immobilized device, a thermoplastic mask, to eliminate set-up errors between each fractions.

All treatment planning was customized using IMRT to obtain high conformity dose distribution. This study used a CTV-PTV uniform margin recipe of 6 mm according to van Herk margin recipe and our clinical protocol. For patients included in this study, prescription dose range was 45 Gy in 25 treatment fractions. For each patient, dose was prescribed to cover 95 % of PTV. A beam arrangement consisting of 7 coplanar beams was used.

IMRT was delivered with 10 MV by a linear accelerator (Elekta Oncology Systems, Crawley, UK). None of these patients received para-aortic treatment.

Set-up errors detection

CBCT was derived prior to treatment fraction and a total number of 105 CBCT image sets were acquired for 10 patients. The selected scan parameter was 125 kVp and 1.6mAs per projection. Automatic rigid volumetric image registration of the CBCT to the planning CT was performed on XVI (Elekta Oncology Systems, Crawley, UK) for the purpose of correcting any misalignment of body position in treatment. Bony match registration method was performed on all bony information available. The registration result reported residual set-up errors in six degrees, three translations T(x, y, z) and three rotations R(θx, θy, θz). Here X was in lateral direction, Y was in anterior-posterior direction, Z was in superior-inferior direction, in treatment machine coordinate.

Equivalent beam parameters

Translational movements can be corrected by shifting treatment couch laterally, longitudinally, and vertically in most treatment machines but rotation movements can only be eliminated by robot treatment couch. However, the robot six-degree couch is not widely equipped. In this study, equivalent beam parameters (gantry, collimator, couch angles) were calculated to compensate the patient treatment position rotation under rigid body assumption. For example, rotation in z-axis can be compensated by inverse rotation of gantry, rotation in y-axis can be compensated by collimator and couch rotation, rotation in x-axis can be eliminated by gantry, couch and collimator simultaneously. An in-house developed program used the algorithm (Yue et al. 2006), was coded to calculate the equivalent gantry, collimator, couch angles under rigid body assumption to compensate the patient treatment position rotation which can not be corrected. With this method, the rotational set-up errors can be equivalently corrected by rotating gantry/collimator/couch while keeping the patient position rather than adjust patient position by using robot treatment couch.

Accumulation of delivered dose

For each patient, errors introduced plans were generated by shifting iso-center position and/or changing gantry, collimator and couch angles (see 2.3). The translational errors of (x, y, z) and rotational errors of (θx, θy, θz) were used to calculate the corresponding errors introduced plans. A total number of 105 errors introduced plans were created for translation only (Tonly) plans and translation plus rotation (TplusR) plans, respectively. The cumulative dose was calculated by multiply a weight equally for each investigated fractions

$$ D_{Cum} = \sum\limits_{n}^{i = 1} {\frac{n}{N}} D_{i} $$

where \( D_{i} \) is dose distribution for \( i \) th fraction. \( n \) is the total number of scanning for one patient,\( N \) is the total number of fraction, here \( N = 25 \).

Dosimetric comparison

The three-dimensional (3D) dose distributions of reference plans (original plans) were compared with two types of errors introduced plans, Tonly plans and TplusR plans. The V100 is defined as the target volume that received 100 % of the prescription dose. All dosimetric parameters V100 for CTV and PTV were compared with relevant values obtained from original plan dose distribution and differences accumulated plan dose distribution for each patient. To estimate the value of rotation corrections, we further compared the differences in the dose distribution obtained with TplusR compared with those obtained with Tonly.

Cervical cancer target always has large cranio-caudal extension. Therefore, rotational uncertainties may potentially cause target displace in particular in the regions furthest away from the treatment center. In this study, targets of CTV and PTV were divided into several sections by distance away from treatment center in Z direction as follows: (1) volume of PTV within 4 cm in Z direction; (2) volume of PTV within 7 cm but larger than 4 cm in Z direction; (3) volume of PTV larger than 7 cm in Z direction. The target dose coverage dosimetric parameter V100 for each CTV and PTV section was calculated to evaluate the dose distribution affected by target displace particularly in the regions furthest away from the treatment center.

Statistics and analysis

The paired-samples t test was used on differences between groups to determine whether correction strategy Tonly and TplusR is valuable, with a threshold of 0.05 for significance. Analyses were performed using SPSS 17.0 statistical software package (SPSS Inc., Chicago, IL, USA).

Results

Translational and rotational set-up errors

Of 105 observations for the 10 patients with cervical cancer, distances from cranial to caudal for CTV and PTV were 19.9 ± 2.3 and 20.9 ± 2.3 cm, respectively. The obtained translational and rotational set-up errors, given in the format of mean ± standard deviation (SD), were X = −0. 6 ± 2.3 mm (range −5.1 to 5.3 mm), Y = −1.6 ± 3.9 mm (range −10 to 8.3 mm) and Z = 2.0 ± 2.4 mm (range −4.7 to 10 mm), RX = −0.2° ± 1.1° (range −2° to 2.3°), RY = −0.4° ± 0.9° (range −2.4° to 1.9°), RZ = −0.1° ± 1° (range −2° to 2.2°). Figure 1a, b show the distribution of translational set-up errors in X, Y, Z axes and rotational set-up errors around X (pitch), Y (roll), Z (yaw) axes as a group of 105 6D registrations for 10 patients with cervical cancer. Figure 2a shows the average and standard deviation of translational difference in 3 axes for each patient. Figure 2b shows the average and standard deviation of rotations around 3 axes for each patient involved in this study.

Fig. 1
figure 1

a Distribution of translational set-up errors in X, Y, Z axes and b rotational set-up errors around X (pitch), Y (roll), Z (yaw) axes for 105 CBCT scans

Fig. 2
figure 2

a The average and standard deviation of translational difference in 3 axes for each patient and b the average and standard deviation of rotations around 3 axes for each patient

Translational shifts in Y direction larger than 6 mm occurred in a considerable number of treatment fractions in 15 of 105 observations. Whereas, translational shifts in X and Z resulted in such shifts were only 0 and 3 fractions, respectively.

Roll and yaw were rarely significant factors in the set-up accuracy as only a single fraction had rotations larger than 2°, respectively. Pitch was the most frequent source of residual rotational errors and resulted in rotation exceeding 2° in 3 fractions.

Target coverage

Total dose deviation of two plans, original plan and error introduced plan, was calculated for each patient. Difference between these two plans were show in Fig. 3. Percentage of V100 for all patients and investigated targets are shown in Table 1. For the original plan and errors introduced plan (TplusR) dose distribution, PTV V100 decreased 3.3 %. Deviations were statistically significant (p < 0.05) (Table 1). However, deviation in V100 % for CTV was not considerable (average 0.3 %, p = 0.03).

Fig. 3
figure 3

Example for total dose deviation of two plans

Table 1 The coverage difference changed of CTV and PTV in the clinic (in percentage)
Table 2 The coverage difference changed of each CTV section in the clinic (in percentage)

For each sections of PTV, V100 obtained with TplusR decreased with increasing distance away from treatment center, as expected. The observed mean differences between TplusR delivery and original plan in PTV > 7 cm, 4 cm < PTV < 7 cm and PTV < 4 cm, as derived from Nucletron Oncentra (Nucletron, Veenendaal, the Netherlands) treatment planning system, were −6.2 % (p = 0.0007), −2.5 % (p = 0.0002) and −0.2 % (p = 0.01), respectively (Table 3). For each section of CTV, differences between original plan and TplusR were hardly noticeable −0.8 % (p = 0.03), −0.1 % (p = 0.44), −0.2 % (p = 0.01) (Table 2).

Table 3 The coverage difference changed of each PTV section in the clinic (in percentage)

Rotation correction

The increased dose coverageΔV100 obtained with TplusR and Tonly were compared in this study. Results show no significantly different when rotational set-up errors were correct before treatment. The results for CTV and PTV were0.08 % (p = 0.14) and 0.60 % (p = 0.02), respectively. Three of ten patients had PTV dose coverage boost larger than 1 % (range 1.1–2.4 %). Only one of ten patients had CTV dose coverage boost of 0.66 % (with PTV dose coverage boost 2.4 %) who suffered two fractions of rotational set-up errors larger than 2° (1 in yaw 2.2° and 1 in pitch 2.3°). The result implying that the effect of rotation correction seems to be very small to improve coverage on CTV and modestly on PTV.

Discussions

Rotational errors were often considered to result in compromised target coverage. Previous margin recipe was determined by applying a common margin recipe according to translational set-up errors only in mostly institutions. It is necessary to evaluate the importance of correct rotational set-up errors when we got the fully scope of 6D set-up errors with CBCT and image registration method. Previous studies were believed have limited value to evaluate impact of dosimetric effects of translational and rotational set-up errors because most of them were neither lack of proper method to reestablish rotational set-up errors nor described the effect of rotational set-up errors based on simulation method but few of them using the clinical-obtained set-up errors. Analysis based on simulation data was not realistic and may deduct to an unreasonable conclusion. In this study, a total number of 105 set-up errors were acquired for 10 patients derived before CBCT image registration and correction. A method to correct the 6D set-up changes by adjusting the gantry, collimator, and couch angles of treatment beams (Yue et al. 2006) was used to accumulate dose by fraction in one CT image and evaluate the dosimetric outcome retrospectively. To our knowledge, this is the first time to evaluate target coverage influenced by combined translational and rotational set-up errors as well as target distance away from the treatment center, for IMRT of cervical cancer using online CBCT images with clinical obtained results.

In this study, significant decreace of PTV dose coverage was observed in TplusP plan compared to original plan. The real volume that received prescription dose of PTV is less than 95 % which is the goal suggested by International Commission on Radiation Units and Measurements. ICRU report 62(Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT): Contents 2010). Further, we investigated whether target distance away from the treatment center affected dose distribution. The result showed that PTV coverage was significantly correlated with distance away from treatment center, which was in accordance with the report (Laursen et al. 2012). But neither combined translational and rotational set-up errors nor target distance away from the treatment center significantly affected the coverage of CTV. Then, a dramatical phenomenon, coverage of PTV did not achieve the goal of 95 % but coverage of all CTVs maintained a level as high as 98 % which surpassed the assumpted goal that miminun dose to CTV was 95 for 90 % of patients when CTV to PTV margin recipes was defined by van Herk, was observed. We speculated reasons for this dramatical phenomenon were: first, a 6 mm CTV to PTV margin was redundant to account for the positioning errors; second, the CTV to PTV margin recipes (2.5 SD of systematic translational set-up errors plus 0.7 SD of random translational set-up errors) would not be appropriate for IMRT of cervical cancer, especially with the scope of 6D. Due to the function of PTV that ensure dose coverage of CTV, a smaller margin maybe achieve the goal that miminun dose to CTV was 95 for 90 % of patients based on tumor control probability (TCP) model, which should be further studied.

Cervical cancer target always has large cranio-caudal extension. Therefore, rotational uncertainties may potentially cause target displace in particular in the regions furthest away from the treatment center. Published study suggested that margin should increase with distance from the isocenter in order to take rotational errors into account (Laursen et al. 2012). In this study, CTV and PTV were divided into several sections to evaluate the necessity of variant margin recipe. The result showed that PTV coverage was significantly correlated with distance away from treatment center, but it does not affect the coverage of CTV. However, it should be awarded that this result only gave advice for cervical cancer patients without para-aortic irradiation with decent errors distribution. For patients with para-aortic irradiation, target volume was longer in superior-inferior direction, which may potentially be more susceptible due to rotational set-up errors (Zhang and Yu 2009).

To further evaluate the impact of rotational set-up errors alone in dosimetry, dose coverage differences obtained with TplusR plans were compared with Tonly plans. By eliminating rotational set-up errors in dose accumulated plan, dose coverage of target increased as expected. However, result showed that the benefit of rotation corrections was insignificant on CTV and modestly on PTV with high patient variation. This result implied that rotation corrections seem to improve target coverage very little, and a PTV margin of 6 mm was sufficient to take into account residual rotational set-up errors for the majority of patients.

It needs to be mentioned that this study did not consider the influece of organ motion, which is a challenge of IMRT for cervical cancer. Therefore, how to intergrate set-up errors (both translational and rotational) and organ motion to determine PTV margin should be further investigated.

Conclusions

For cervical cancer IMRT, the dosimetric influence for CTV of the distance away from treatment center and the translational and rotational set-up errors, not corrected after CBCT scanning and image registration was small. A PTV margin of 6 mm was sufficient to take into account geometry uncertainties for most patients. A smaller margin should be further studied with the scope of 6D.

References

  • Ahamad A, D’Souza W, Salehpour M, Iyer R, Tucker SL, Jhingran A, Eifel PJ (2005) Intensity-modulated radiation therapy after hysterectomy: comparison with conventional treatment and sensitivity of the normal-tissue-sparing effect to margin size. Int J Radiat Oncol Biol Phys 62(4):1117–1124. doi:10.1016/j.ijrobp.2004.12.029

    Article  Google Scholar 

  • Antolak JA, Rosen II, Childress CH, Zagars GK, Pollack A (1998) Prostate target volume variations during a course of radiotherapy. Int J Radiat Oncol Biol Phys 42(3):661–672

    Article  Google Scholar 

  • Beard CJ, Kijewski P, Bussiere M, Gelman R, Gladstone D, Shaffer K, Plunkett M, Castello P, Coleman CN (1996) Analysis of prostate and seminal vesicle motion: implications for treatment planning. Int J Radiat Oncol Biol Phys 34(2):451–458

    Article  Google Scholar 

  • Brixey CJ, Roeske JC, Lujan AE, Yamada SD, Rotmensch J, Mundt AJ (2002) Impact of intensity-modulated radiotherapy on acute hematologic toxicity in women with gynecologic malignancies. Int J Radiat Oncol Biol Phys 54(5):1388–1396

    Article  Google Scholar 

  • Chen MF, Tseng CJ, Tseng CC, Kuo YC, Yu CY, Chen WC (2007) Clinical outcome in posthysterectomy cervical cancer patients treated with concurrent Cisplatin and intensity-modulated pelvic radiotherapy: comparison with conventional radiotherapy. Int J Radiat Oncol Biol Phys 67(5):1438–1444. doi:10.1016/j.ijrobp.2006.11.005

    Article  Google Scholar 

  • Guckenberger M, Meyer J, Vordermark D, Baier K, Wilbert J, Flentje M (2006) Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study. Int J Radiat Oncol Biol Phys 65(3):934–942. doi:10.1016/j.ijrobp.2006.02.019

    Article  Google Scholar 

  • Kaiser A, Schultheiss TE, Wong JY, Smith DD, Han C, Vora NL, Pezner RD, Chen YJ, Radany EH (2006) Pitch, roll, and yaw variations in patient positioning. Int J Radiat Oncol Biol Phys 66(3):949–955. doi:10.1016/j.ijrobp.2006.05.055

    Article  Google Scholar 

  • Kim GY, Pawlicki T, Le QT, Luxton G (2008) Linac-based on-board imaging feasibility and the dosimetric consequences of head roll in head-and-neck IMRT plans. Med Dosim 33(1):93–99. doi:10.1016/j.meddos.2007.05.004

    Article  Google Scholar 

  • Laursen LV, Elstrom UV, Vestergaard A, Muren LP, Petersen JB, Lindegaard JC, Grau C, Tanderup K (2012) Residual rotational set-up errors after daily cone-beam CT image guided radiotherapy of locally advanced cervical cancer. Radiother Oncol 105(2):220–225. doi:10.1016/j.radonc.2012.08.012

    Article  Google Scholar 

  • Letourneau D, Wong JW, Oldham M, Gulam M, Watt L, Jaffray DA, Siewerdsen JH, Martinez AA (2005) Cone-beam-CT guided radiation therapy: technical implementation. Radiother Oncol 75(3):279–286. doi:10.1016/j.radonc.2005.03.001

    Article  Google Scholar 

  • Lim K, Kelly V, Stewart J, Xie J, Cho YB, Moseley J, Brock K, Fyles A, Lundin A, Rehbinder H, Milosevic M (2009) Pelvic radiotherapy for cancer of the cervix: is what you plan actually what you deliver? Int J Radiat Oncol Biol Phys 74(1):304–312. doi:10.1016/j.ijrobp.2008.12.043

    Article  Google Scholar 

  • Lips IM, van der Heide UA, Kotte AN, van Vulpen M, Bel A (2009) Effect of translational and rotational errors on complex dose distributions with off-line and on-line position verification. Int J Radiat Oncol Biol Phys 74(5):1600–1608. doi:10.1016/j.ijrobp.2009.02.056

    Article  Google Scholar 

  • Mackie TR, Kapatoes J, Ruchala K, Lu W, Wu C, Olivera G, Forrest L, Tome W, Welsh J, Jeraj R, Harari P, Reckwerdt P, Paliwal B, Ritter M, Keller H, Fowler J, Mehta M (2003) Image guidance for precise conformal radiotherapy. Int J Radiat Oncol Biol Phys 56(1):89–105

    Article  Google Scholar 

  • Peng J, Chang X, Xu Z, Wang J, Guan X, Cheng L, Zhang Z, Hu W (2013) Residual rotational error of X-ray volume imaging system and its dosimetric effect in spinal tumor radiation therapy. Aust Phys Eng Sci Med 36(2):193–199. doi:10.1007/s13246-013-0194-0

    Article  Google Scholar 

  • Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT): Contents (2010). J ICRU 10 (1):NP. doi:10.1093/jicru/ndq002

  • Quint S, de Boer HC, van Sornsen de Koste JR, Heijmen BJ, Olofsen-van Acht MJ (2001) Set-up verification of cervix cancer patients treated with long treatment fields; implications of a non-rigid bony anatomy. Radiother Oncol 60(1):25–29

    Article  Google Scholar 

  • van de Bunt L, van der Heide UA, Ketelaars M, de Kort GA, Jurgenliemk-Schulz IM (2006) Conventional, conformal, and intensity-modulated radiation therapy treatment planning of external beam radiotherapy for cervical cancer: the impact of tumor regression. Int J Radiat Oncol Biol Phys 64(1):189–196. doi:10.1016/j.ijrobp.2005.04.025

    Article  Google Scholar 

  • van Herk M (2004) Errors and margins in radiotherapy. Semin Radiat Oncol 14(1):52–64. doi:10.1053/j.semradonc.2003.10.003

    Article  Google Scholar 

  • Verellen D, Soete G, Linthout N, Van Acker S, De Roover P, Vinh-Hung V, Van de Steene J, Storme G (2003) Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging. Radiother Oncol 67(1):129–141

    Article  Google Scholar 

  • Yue NJ, Knisely JP, Song H, Nath R (2006) A method to implement full six-degree target shift corrections for rigid body in image-guided radiotherapy. Med Phys 33(1):21–31

    Article  Google Scholar 

  • Zeidan OA, Langen KM, Meeks SL, Manon RR, Wagner TH, Willoughby TR, Jenkins DW, Kupelian PA (2007) Evaluation of image-guidance protocols in the treatment of head and neck cancers. Int J Radiat Oncol Biol Phys 67(3):670–677. doi:10.1016/j.ijrobp.2006.09.040

    Article  Google Scholar 

  • Zhang X, Yu H (2009) Evaluation of pelvic lymph node coverage of conventional radiotherapy fields based on bony landmarks in Chinese cervical cancer patients using CT simulation. J Zhejiang Univ Sci B 10(9):683–688. doi:10.1631/jzus.B0920114

    Article  Google Scholar 

Download references

Authors’ contributions

BW conceived the study idea; XZ, GS, JL, BW participated in the data collection; BW undertook the data analysis; and XZ, BW contributed to compiling the manuscript and have approved the final version. All authors read and approved the final manuscript.

Acknowledgements

The authors would like to thank the colleges who volunteered their time to participate in the study.

Competing interests

The authors declare they have no competing interests.

Ethical approval

All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the Helsinki Declaration of 1975, as revised in 2008 (5). Informed consent was obtained from all patients for being included in the study. This study was approved by Zhejiang Cancer Hospital institutional review board.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Bin-bing Wang.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, X., Shan, Gp., Liu, Jp. et al. Margin evaluation of translational and rotational set-up errors in intensity modulated radiotherapy for cervical cancer. SpringerPlus 5, 153 (2016). https://doi.org/10.1186/s40064-016-1796-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40064-016-1796-2

Keywords