Abstract

This study aimed to develop a modified Tessari method for producing more sclerosing foam in treatment of extensive venous malformations. Sclerosing foam was produced by using Tessari method and the modified Tessari method. The procedure of the later was as follows: prepared foam in a sclerosant–air ratio of 1:4; connected three disposable 10 ml syringes to two medical three-way taps; drawn 4 ml of liquid sclerosant into one syringe and 16 ml averagely of air into the other two; then moved the plungers of all syringes back and forth for 20 times to produce sclerosing foam. The volume and foam half time (FHT) of foam produced by the two methods were compared. The average volume of sclerosing foam produced by Tessari method and the modified Tessari method were 9.8 and 19.7 ml, and assessed to have statistical difference. The FHT of foam produced by the two methods were 120 and 150 s, and assessed to have statistical difference. In conclusion, the modified Tessari method could produce more fresh and stable sclerosing foam.

Keywords

Sclerosing foam Tessari technique Modified Tessari method Sclerotherapy Venous malformations

Background

During the treatment of venous malformations (VMs) using foam sclerotherapy, one of the most popular methods to produce foam was the Tessari method which used pumping cycles of liquid and air in-and-out of a double syringe system (Nastasa et al. 2015; Wolmann 2010; Tessari 2001, 2005). In clinical treatment, sclerosing foam was prepared usually by using two 10 ml syringes; however, no more than 10 ml of sclerosing foam could be produced at a time in this way. Tessari method was originally used in the treatment of varicose veins, in most cases of which 10 ml of fresh sclerosing foam might be sufficient (Tessari et al. 2001). However, this volume might be insufficient in treating some VMs, especially when the region of lesion was extensive. Sufficient fresh sclerosing foam was needed for the treatment of extensive VMs.

There were several ways to get more fresh sclerosing foam, for example, producing foam once more, using bigger syringes or more operators and devices to produce simultaneously. However, all these methods had more or less defects. Here, we reported a modified Tessari method by using three 10 ml syringes which could produce more fresh foam at a time; the volume and the stability of foam produced by the two methods were compared. Foam half time (FHT), the time when half of the original liquid sclerosant was reverted to liquid state, was adopted to assess the stability of the foam.

Methods

Disposable 10 ml syringes, medical three-way taps and 1 % lauromacrogol injection (Shanxi Tianyu Pharmaceutical Co., Ltd.) were used in this study.

Procedure of Tessari method (Tessari et al. 2001): drawn 2 ml of liquid sclerosant into one 10 ml syringe and 8 ml of air into the other 10 ml syringe; connected both syringes to a medical three-way tap; then, moved the plungers of both syringes back and forth for 20 times to produce sclerosing foam (Fig. 1).
Fig. 1

Tessari method

Procedure of the modified Tessari method: drawn 4 ml liquid sclerosant into a 10 ml syringe and 16 ml air averagely into the other two 10 ml syringes; then connected all syringes to two parallel medical three-way taps; then moved the plungers of all syringes back and forth for 20 times to produce sclerosing foam (Fig. 2).
Fig. 2

The modified Tessari method

Detection of the foam volume and the foam FHT: Once the foam was produced, it was immediately pushed into 10 and 20 ml scalar syringes respectively, the scalar syringes were placed vertically with the piston beneath. Macroscopic appearance of the foam was observed and volume of the foam was recorded. The foam in the syringes gradually separated into liquid and air, with the liquid sclerosant gathered at the bottom, when half volume of the liquid sclerosant was reverted, the time as FHT was recorded.

The detection above was measured respectively for 10 times by the same operator. Each foam was prepared by using new sclerosant and all experiments were performed at an ambient temperature of 25 °C. Experimental values were compared by an independent-samples t test and analysed using SPSS 19.0. A p value ≤0.05 was considered significant.

Results

Right after production, the bubbles in foam were small and homogenous, when the foam began to re-convert to a liquid and air separated mix, foam bubbles got bigger and inhomogeneous, eventually foam converted to two components: liquid and air.

The average volume of sclerosing foam produced by Tessari method and the modified Tessari method were 9.8 and 19.7 ml, and assessed to have statistical difference.

The FHT of foam produced by Tessari method and the modified Tessari method were 120 and 150 s, and assessed to have statistical difference (Table 1).
Table 1

Average volume and foam half time (FHT) of foam produced by Tessari method and modified Tessari method

 

Tessari method

The modified Tessari method

The average volume (ml)a

9.8 ± 0.2

19.7 ± 0.3

The FHT (s)a

120 ± 4.13

156 ± 6.85

aThe average volume and FHT were expressed as mean values ± standard deviation

Discussion

Owing to its high efficiency, safety and economical advantages, foam sclerotherapy was widely used in treatment of vascular abnormalities (Luebke and Brunkwall 2008; Myers and Roberts 2009; Park et al. 2014; Van der Vleuten et al. 2014; Wolmann 2010). The high efficiency of foam sclerotherapy was closely related to the physical obstruction of blood flow and the displacing the intravascular blood, which extended extent of contaction between the sclerosants and the vessel wall (Ikponmwosa et al. 2010; Wolmann 2010; Eckmann 2009; Van Deurzen et al. 2011). Therefore, the therapeutic effects related to whether the cavity of VMs could be replete by fresh sclerosing foam; In our clinical work, sometime more than 10 ml fresh sclerosing foam was needed for extensive lesion, some authors found that 10 ml or more sclerosing foam was safety and efficacy, and no serious complications were observed (Stimpson et al. 2012; Chen et al. 2015). In our experience, the most common complications were temporal pain with injection, swelling after injection and skin ulcer in some patients.

In our experimental and clinical experience, it was inconvenient to use 20 ml syringes to produce foam; and other authors had reported that 10 ml syringes were the best choice to produce foam considering normal handling and foam stability (Wolmann 2010; Nastasa et al. 2015). Tessari method using two 10 ml syringes had become one of the most popular ways to produce foam. In clinical treatment, no more than 10 ml of foam could be produced at a time by using two 10 ml syringes, which might be insufficient in treating some extensive VMs.

To get more foam at a time, we might repeat the producing procedure once more or add more operators and devices. But the process of injecting foam into extensive VMs might be interrupted. During the interval between two adjacent injections, the previous foam would be washed away by the blood. Therefore, injecting sufficient foam at a time into the cavity of extensive VMs produced better therapeutic effects (Van Deurzen et al. 2011). Furthermore, there was another disadvantage when using the same syringe to prepare foam for the second time, it would be hard for the plunger to slide smoothly during preparation and injection. Adding more operators and devices to produce foam simultaneously meant to use more human and material resources. The modified Tessari method could produce about 20 ml fresh foam at a time which, in our clinical experience, would be sufficient in treating some of extensive VMs.

FHT was a key and usually used parameter of the stability of sclerosing foam. In this study, the FHT of sclerosing foam produced by the modified Tessari method was longer than that by Tessari method, showing that the sclerosing foam produced by the modified Tessari method was more stable.

Conclusions

The modified Tessari method could produce more fresh and stable sclerosing foam at a time which was profitable in treating extensive VMs.

Abbreviations

VMs: 

venous malformations

FHT: 

foam half time

Declarations

Authors’ contributions

All authors read and approved the final manuscript.

Acknowledgements

This work was supported by the National Natural Science Foundation of China (No. 81172572), the Key Research and Development Project of Shandong Province (No. 2015GGH318012) and Basic Research of Qilu Hospital (2014QLKY).

Competing interests

The authors declare that they have no competing interests.

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Authors’ Affiliations

(1)
School of Stomatology, Shandong University
(2)
Department of Oral and Maxillofacial Surgery, Stomatology Hospital of Jinan
(3)
Department of Oral and Maxillofacial Surgery and Institute of Dental Medicine, Qilu Hospital, Shandong University
(4)
Department and Institute of Dental Medicine, Qilu Hospital, Shandong University

References

  1. Chen AW, Liu YR, Li K, Wang T, Liu SH (2015) Efficacy of sclerotherapy with radio-opaque foam guided by digital subtraction angiography for the treatment of complex venous malformations of the head and neck. Br J Oral Maxillofac Surg 53(9):809–813View ArticleGoogle Scholar
  2. Eckmann DM (2009) Polidocanol for endovenous microfoam sclerosant therapy. Expert Opin Investig Drugs 18:1919–1927View ArticleGoogle Scholar
  3. Ikponmwosa A, Abbott C, Graham A, Homer-Vanniasinkam S, Gough MJ (2010) The impact of different concentrations of sodium tetradecyl sulphate and initial balloon denudation on endothelial cell loss and tunica media injury in a model of foam sclerotherapy. Eur J Vasc Endovasc Surg 39:366–371View ArticleGoogle Scholar
  4. Luebke T, Brunkwall J (2008) Systematic review and metaanalysis of endovenous radiofrequency obliteration, endovenous laser therapy, and foam sclerotherapy for primary varicosis. J Cardiovasc Surg (Torino) 49:213–233Google Scholar
  5. Myers KA, Roberts S (2009) Evaluation of published reports of foam sclerotherapy: what do we know conclusively? Phlebology 24(6):275–280View ArticleGoogle Scholar
  6. Nastasa V, Samaras K, Ampatzidis Ch, Karapantsios TD, Trelles MA, Moreno-Moraga J, Smarandache A, Pascu ML (2015) Properties of polidocanol foam in view of its use in sclerotherapy. Int J Pharm 478:588–596View ArticleGoogle Scholar
  7. Park Sang Woo, Yun Ik Jin, Hwang Jae Joon, Lee Song Am, Kim Jun Seok, Chee Hyun Keun, Chang Il Soo (2014) Fluoroscopy-guided endovenous sclerotherapy using a microcatheter prior to endovenous laser ablation: comparison between liquid and foam sclerotherapy for varicose tributaries. Korean J Radiol 15(4):481–487View ArticleGoogle Scholar
  8. Stimpson P, Hewitt R, Barnacle A, Roebuck DJ, Hartley B (2012) Sodium tetradecyl sulphate sclerotherapy for treating venous malformations of the oral and pharyngeal regions in children. Int J Pediatr Otorhinolaryngol 76(4):569–573View ArticleGoogle Scholar
  9. Tessari L (2001) Extemporary sclerosing foam according to personal method: experimental clinical data and catheter usage. Int Angiol Suppl 1:54Google Scholar
  10. Tessari L (2005) Nouvelle technique d’obtention de la scle´ro-mousse. Phlébologie 53:129Google Scholar
  11. Tessari L, Cavezzi A, Frullini A (2001) Preliminary experience with a new sclerosing foam in the treatment of varicose veins. Dermatol Surg 27:58–60Google Scholar
  12. Van der Vleuten CJ, Kater A, Wijnen MH, Schultze Kool LJ, Rovers MM (2014) Effectiveness of sclerotherapy, surgery, and laser therapy in patients with venous malformations: a systematic review. Cardiovasc Intervent Radiol 37:977–989Google Scholar
  13. Van Deurzen B, Ceulen RP, Tellings SS, van der Geld C, Nijsten T (2011) Polidocanol concentration and time affect the properties of foam used for sclerotherapy. Dermatol Surg 37:1448–1455View ArticleGoogle Scholar
  14. Wolmann JC (2010) Sclerosant foams Stabilities, physical properties and rheological behavior. Phlebologie 39:208–217Google Scholar

Copyright

© Xu et al. 2016