From: Improvement of the detection efficiency calibration and homogeneity measurement of Si-SPAD detectors
Uncertainty components | Uncertainty (%) |
---|---|
Planck constant, h | 2.52 × 10−7 |
Speed of light, c | 0.0 |
Wavelength, λ | 0.0075 |
Amplification factor, A 1 | 0.0021 |
Amplification factor, A 2 | 2.08 × 10−6 |
Amplification factor, A 3 | 2.08 × 10−6 |
Ratio V 1/V Mon1, Q 1 | 0.004 |
Ratio V 2/V Mon2, Q 2 | 0.015 |
Ratio V 3/V Mon3, Q 3 | 0.05 |
Ratio CR/V MonSPAD, Q 4 | 0.036 |
Spectral responsivity of integrating sphere with Si-diode, s Si | 0.15 |
Factor for the use of two filters, F filt | 0.05 |
Combined uncertainty, u c | 0.162 |