Table 3 Image noise and radiation dose at different tube voltages in the case of CNR at 14

From: Radiation dose reduction based on CNR index with low-tube voltage scan for pediatric CT scan: experimental study using anthropomorphic phantoms

Tube voltage (kVp)

Tube current (mA)

Surface dose (mGy)

Center dose (mGy)

Measurd image noise (HU)

New born

 FBP

  80

20

0.6

0.4

37

  100

15

0.6

0.6

30

  120

15

0.9

0.9

25

 IR

  80

20

0.6

0.5

37

  100

15

0.7

0.6

30

  120

15

0.9

0.9

25

1 year old

 FBP

  80

45

1.1

0.9

35

  100

30

1.3

1.1

29

  120

25

1.5

1.5

25

 IR

  80

40

0.9

0.7

37

  100

25

1.1

1.0

30

  120

20

1.4

1.2

25