Table 6 The statistics of 66 clinical throat swab samples

From: Genotyping and detection of common avian and human origin-influenza viruses using a portable chemiluminescence imaging microarray

Microarray Real-time PCR
Positive Negative
Positive (genotype) 51 (30) 4 (0)
Negative (genotype) 5 (0) 6 (0)
  1. The sensitivity of microarray compared to real-time PCR was 91.1 %
  2. The specificity of microarray compared to real-time PCR was 60.0 %
  3. The positive predictive values of microarray compared to real-time PCR was 92.7 %
  4. The negative predicative values of microarray compared to real-time PCR was 54.5 %
  5. The diagnose accordance rate of microarray compared to real-time PCR was 86.4 %
  6. The percent of genotyping for positive sample of microarray was 53.6 %