The outcomes of patients with very small coronary artery disease treated with thin strut cobalt chromium bare metal stents: an observational study

Abstract

Background

Percutaneous coronary interventions (PCI) in coronary artery disease (CAD) with very small vessel diameters remains controversial and challenging. These lesions are usually more diffuse, calcified and tortuous. The usage of thin strut bare metal stents (BMS) with excellent crossing profiles in a very small caliber coronary lesions has increased the likelihood of procedural success.

Objectives

This observational study assessed the 9-month clinical outcomes in an ‘all-comers’ population with very small caliber CAD after implantation of thin strut cobalt chromium BMS.

Methods

Thin strut cobalt chromium BMS implantation in a priori pre-defined subgroups was investigated in a non-randomized, international, multi-center ‘all-comers’ observational study. Primary end-point was the 9-month clinically driven target lesion revascularization (TLR) rate. Secondary end-points included the 9-month major adverse cardiac event (MACE) and procedural success rates. Data collection was done using an established electronic data acquisition form with built-in plausibility checks.

Results

A total of 783 patients with a mean age of 70.4 ± 12.8 years were enrolled, 205 (26.2 %) of them had vessel diameters of 2.5 mm and smaller which was defined as CAD with very small reference vessel calibers. Older age and diabetics were associated with higher incidences of very small caliber vessels. The mean reference vessel diameter in the very small vessel group was 2.05 ± 0.27 mm and mean diameter for vessels >2.5 mm was 3.41 ± 0.55 mm. Pre-dilatation was performed more often in the very small vessel patients (52.2 vs. 42.2 %; p value 0.007). There was no difference in the overall technical success rates in very small vessel disease group (97.9 vs. 97.7 %). The 9-month TLR rate was 6.3 % for the very small vessels and 3.7 % for vessels >2.5 mm (p = 0.129). The 9-month and in-hospital MACE rates in the very small vessel group and patient with vessel diameters >2.5 mm were not significantly different (13.1 vs. 9.2 %; p = 0.1265 and 5.2 vs. 3.7 %; p = 0.349) respectively.

Conclusion

This study has demonstrated that the use of thin strut cobalt chromium BMS in very small vessel CAD was reasonably safe and efficacious in the context of ‘real-world’ practice.

Background

Treatment for very small vessel atherosclerosis with revascularization procedures remains a challenge in daily clinical practice. Coronary artery bypass grafting (CABG) in this group of patients is limited by high rates of graft failure (O’Connor et al. 1996) whereas percutaneous coronary interventions (PCI) are associated with increased risks of restenosis and adverse clinical outcomes (Elezi et al. 1998). PCI with bare metal stent (BMS) as compared to plain old balloon angioplasty (POBA) in small vessel disease only revealed conflicting results and only modest superiority of BMS over POBA (Park et al. 2000; Kastrati et al. 2000; Koning et al. 2001; Agostoni et al. 2005). BMS are generally associated with a higher risk of in-stent restenosis (ISR). The advances of drug-eluting stent (DES) technology have reduced the risk of in-stent restenosis (ISR) tremendously (Morton et al. 2003; Kirtane et al. 2009; Park et al. 2003; Morice et al. 2002; Moses et al. 2003). However, the incorporation of various drugs on the DES platform lead to an effective strut thickness increase. Hence, the usage of DES in very small vessel may still pose some potential issue including flexibility, deliverability and a continued risk of late and very late stent thrombosis (Farooq et al. 2011; Pfisterer et al. 2006). ISAR-STEREO and ISAR-STEREO 2 trials have demonstrated that thinner strut devices are associated with a significant reduction of angiographic restenosis after coronary artery stenting (Kastrati et al. 2001; Pache et al. 2003). Usage of a novel thin strut cobalt-chromium BMS (Coroflex® Blue) in a ‘real-world’ setting including patients with acute myocardial infarction was proven to be safe and efficacious. Event free survival in this cohort of patients at 6 months was 90.8 % and the TLR rate for ISR at 6 months was 5.5 %. Furthermore, the cumulative stent thrombosis rate of 1.6 % and the absence of late stent thrombosis within 6 months is remarkably low for an all-comer cohort (Bocksch et al. 2010).

Currently, BMS is still used in cases of large luminal diameters, acute STEMI, in patients with high bleeding risk, anticipated surgical procedures for underlying illness and in situations where financial limitations may be an issue. With the latest stent architecture developments in BMS, there is a question whether it is still practical or even safe to implant BMS in lesions with very small vessel diameters.

The definition of small vessel coronary arteries involves more than an arbitrary upper limit of lumen diameters between 2.5 and 2.75 mm. Coronary stenting in small vessels has been shown to be safe, feasible and effective in the long term (Pache et al. 2006). With the availability of 2.25 and 2.0 mm diameter stents, the term very small vessel disease was suggested to those that are amenable to PCI with a 2.25 and 2.0 mm device. Despite varying definitions in the literature, small vessel coronary artery disease (CAD) is highly prevalent. It is estimated that approximately 20–30 % of patients with symptomatic CAD have small vessel CAD (Biondi-Zoccai et al. 2007). However, coronary stenting in very small vessels (lumen diameter ≤ 2.5 mm) remain controversial and challenging.

The objective of this registry was to document the safety and efficacy of thin strut BMS in an unselected patient population in which BMS implantation was deemed reasonable by the treating physician. This observational study assessed the 9-month clinical outcomes in an ‘all comers’ population with a focus lesions with very small coronary arteries in a ‘real-world’ setting.

Methods

Thin strut bare metal stenting in a priori defined subgroups was investigated. This was a non-randomized, international, multi-center ‘all-comers’ observational study designed to evaluate the safety and efficacy of the thin strut Coroflex® Blue Neo/Ultra coronary stent system (B.Braun Vascular Systems, Berlin, Germany). Patients were prospectively enrolled in 3 Asian and 17 European centers. In France this study was approved by the “Comité Consultative sur le Traitement de I’Information en matière de Recherche dans le domaine de la Santé (CTIRS dossier no. 12.384) and the Commission Nationale de I’informatique et des Libertés (CNIL, demande d’autorisation no. 912431). In Malaysia, this study received approval from the University of Malaya Medical Centre ethics committee, the Universiti Kebangsaan Malaysia ethics committee as well as from National Medical Research and Ethics Committee. The legal requirements are different in the participating countries. Inasmuch this clinical assessment was an observational study under routine use, the conduct in Germany was regulated by the German Device law (Medizin Produkte Gesetz §23b) which did not require a formal ethics vote at the time this registry was conducted. Likewise in the Netherlands, Switzerland and Croatia, ethics votes were not required given that data collection was done in a pseudonymized manner (no patient initials, only birth years and not birth dates). Also these patients would have been treated with the device independent of the data collection. The clinical follow-up was part of the routine patient care established in each participating country. With the exception of France, two votes were mandatory for the collection of data and one for the ethical conduct of the study (please see the attached documents). The responsibility for the documentation of the local requirements was at the individual centers. The Declaration of Helsinki was respected in all countries. An informed consent text specific to the registry approved by the local ethics committees was to be signed by patients prior to their enrolment as mandated by the local ethics committees.

Patients aged ≥18 years with stable angina or objective proof of ischemia or patients with acute coronary syndrome (ACS) had to meet the requirements for PCI (Windecker et al. 2014). The decision to implant BMS solely depended on the operator preference and experience. Single or multiple vessel stenting was allowed in de novo or re-stenotic lesions. The adjunct treatment with drug coating balloons (DCB) was left to the discretion of the operator. A maximum lesion length of 28 mm and reference vessel diameters from 2.0 to 4.0 mm were permissible. Vascular access via the femoral and radial route was permitted with recommended introducer sheaths of at least 5 French in diameter. Operators could also choose direct stenting at their discretion. A standard bolus of intravenous heparin (70 IU/kg) was given in all patients and supplemented when required. Platelet aggregation inhibitor loading was initiated prior to the procedure.

The cobalt chromium Coroflex® Blue Ultra/Neo is one of the latest generation coronary BMS characterized by its ultra-low strut thickness of 60 μm (Coroflex® Blue Neo, vessel diameters from 2.75 to 4.0 mm) and of 50 μm (Coroflex® Blue Ultra, vessel diameters from 2.0 to 2.5 mm). The stents are made of implantable, high-grade, surgical CoCr alloy (L605, ISO 5832-5 or ASTM F.90) with an established biocompatibility. This alloy has been successfully used as a material for stents and other surgical implants for many years. The stents were designed to improved deliverability in lesions that are extremely difficult to cross while reducing the metal volume in the treated vessel. The smaller caliber version is made up of units with 6 crowns whereas the larger version displays 9 crowns. Figure 1 showed the illustration of the stent. The device is available in lengths of 8–32 mm with a crossing profile of less than 0.85 mm.

Fig. 1
figure 1

Coroflex® Blue Ultra stent design (top panel). Coroflex® Blue Ultra is made up of units with 6 crowns, Coroflex® Blue Neo with 9 crowns, crimped stent (bottom panel)

DES implantation is the standard of care for most indications; however, the use of BMS still has a role in selected treatment populations particularly in some countries where the device costs are covered by the patient. Furthermore, in patients who cannot tolerate dual antiplatelet therapy (DAPT) for long periods or patients with difficult vessel anatomies. The decision to use the thin strut BMS was based on the patient’s clinical presentation and on the physician’s preferences (Windecker et al. 2014). The used investigational devices are one of the many BMS still being used in ‘real-world’ practice. The preferred BMS in this observational study has the advantage of having thin struts to allow a very favorable lesion crossability in difficult to treat lesions as compared to other less deliverable BMS.

At the discretion of the treating physician clopidogrel 75 mg/day, prasugrel 10 mg/day or ticagrelor 2 × 90 mg/day were prescribed at least 1 month post procedure and acetylsalicylic acid 100–325 mg/d for life.

In this sub group analysis we are comparing patients with very small vessel ≤2.5 mm diameter and vessel >2.5 mm diameter. From the previous literature, the definition of small vessel coronary arteries varies between 2.5 and 2.75 mm (Pache et al. 2006; Biondi-Zoccai et al. 2010). With the introduction of 2.0 and 2.25 mm diameter stents, any vessel feasible for interventions with 2.0 and 2.25 mm devices were defined as very small caliber CAD.

Endpoints and definitions

Clinically driven target lesion revascularization (TLR) at 9-month post procedure as the composite of re-PCI or CABG was defined as the primary endpoint. Secondary endpoints included major adverse cardiac events (MACE) as the ensemble of TLR, cardiac death and myocardial infarction (MI). Rates of definite acute/sub-acute stent thrombosis and procedural success rates were also documented. MI was diagnosed with corresponding ECG changes and/or elevated troponin T or troponin I. Definite acute/sub-acute stent thrombosis were based on the ARC criteria (Cutlip et al. 2007). Cardiac death was defined as death from all causes except when the cause of death was proven non-cardiac. Procedural success was defined as achievement of less than 30 % residual in-segment percent diameter stenosis and TIMI flow 3 after the procedure using the assigned device only.

Data collection

All data were collected through a web based data acquisition system previously used in other observational studies (Cutlip et al. 2007; Wöhrle et al. 2012; Zeymer et al. 2014). National principal investigators were responsible for the accuracy of their datasets and performed source data verification when the routinely performed web based plausibility checks indicated discrepancies.

Statistical analysis

To prove Gaussian distribution, the Kolmogorov–Smirnov-test was used to justify the calculation of the mean and standard deviation unless otherwise indicated. Dichotomous variables were evaluated using the chi2-test while continuous variables were compared with the unpaired two-tailed student´s t test. Non-parametric test such as the Wilcoxon–Mann–Whitney rank test were used whenever applicable. For all tests the significance level α was 0.05. Based on prior result with a predecessor device (Bocksch et al. 2010), a literature value of 5.5 % for TLR was chosen. With a performance goal margin of +4.0 % and an expected 9-month TLR rate of 6.0 % a total of 490 patients would be needed to detect a difference between the null hypothesis proportion and the alternative proportion. A logistic regression model for 9-month MACE was implemented for defined cardiovascular risk factors such as presence of diabetes or ACS and lesion morphological variables such as vessel diameter, lesion length and presence of calcification. Statistical analyses were done with SPSS version 20.0 (IBM, Munich, Germany). The biometric estimate was calculated with nQuery/nTerim version 2.0 (Statistical Solutions Ltd. Cork, Ireland).

Results

Baseline demographics

A total of 783 patients underwent BMS implantations between 1.1.2012 and 31.12.2013. Of these patients, 205 (26.2 %) have very small vessel diameters (≤2.5 mm) and 578 (73.8 %) have vessel diameters >2.5 mm (Table 1). Patients with very small vessel diameters were older (mean age of 72.5 ± 11.3 years vs 69.7 ± 13.1 years) more diabetes (31.7 vs 22.8 %) compared to those with vessel diameters >2.5 mm. More women (24.4 vs 20.9 %; p = 0.303) and end stage renal disease patients (7.3 vs 4.0 %; p = 0.056) were in the very small vessel group. Remarkably, 316 (40.4 %) of the overall patients presented with ACS, 42.0 % in very small vessel versus 39.8 % in vessel diameters >2.5 mm. From the total of 20.8 % who had STEMI, 22.4 % with very small vessel versus 20.2 % with vessel diameters >2.5 mm.

Table 1 Patient demographics

Angiographic characteristics, procedural data and co-medication

Baseline angiographic characteristics and procedural data are shown in Table 2. There was a total of 880 coronary lesions which were treated during this study. 247 (28.1 %) lesions treated were in very small vessels primarily in the left anterior descending artery (39.7 %) and left circumflex artery (37.2 %). Of note, 40.1 % of the total lesions were complex (type B2/C) according to the modified American College of Cardiology (ACC) and American Heart Association (AHA) classification (Ryan et al. 1998). Smaller vessel lesions trended to be more diffusely diseased with higher calcium burden. Pre-dilatation was performed more often in the very small vessels (52.2 %) as compared to vessel diameters >2.5 mm (42.2 %). The mean reference diameter for very small vessel group was 2.05 ± 0.27 mm; whereas the mean reference diameter for vessel diameters >2.5 mm was 3.41 ± 0.55 mm. There was no difference in terms of stent balloon inflation pressure (very small vessel 14.8 ± 3.1 atm vs. vessel diameters >2.5 mm 15.1 ± 3.0 atm) and the overall technical success rate in very small vessel group was 97.9 versus 97.7 % in the patient group with vessel diameters >2.5 mm. The new oral antiplatelet drugs prasugrel/aspirin and ticagrelor/aspirin were administered more frequently in the very small vessel (11.2 vs 9.5 %) and (7.8 vs 2.4 %) compared to vessel diameters >2.5 mm respectively (Table 3). The duration of dual antiplatelet therapy (DAPT) between the two groups were not statistically different.

Table 2 Lesion characteristics and procedural data
Table 3 Peri-procedural drug therapy

Clinical outcomes

Overall follow-up at 9 months was 93.9 %. The 9-month TLR rate was 6.3 % for very small vessels and 3.7 % for vessels >2.5 mm (p = 0.129). Figure 2 showed the Kaplan–Meier curve for freedom from TLR. The log-rank test revealed no statistical difference with p = 0.112. 9-month and in-hospital MACE between very small vessels and vessels >2.5 mm were not significantly different either (13.1 vs 9.2 %; p = 0.1265 and 5.2 vs 3.7 %; p = 0.349) respectively. There was no case of acute stent thrombosis observed in either group (Table 4).

Fig. 2
figure 2

Kaplan–Meier curve for freedom from TLR in very small vessels versus larger vessels

Discussion

Vessel size remains an important determinant of adverse outcomes even in the DES era (Togni et al. 2007). While DES are consistently less prone to restenosis, efforts are being made to narrow the efficacy gap between BMS and DES. Clinical studies have shown the potential advantage of thinner stent struts in reducing the restenosis rate (Suttorp et al. 2015). Although DES are preferred as a default treatment strategy, BMS are still implanted for reasons such as limited financial resources in some Asian countries or contraindications to prolonged DAPT. There is limited data on PCI in very small vessels involving ‘real-world’ patients. The Arthos Pico Austrian Multi-center registry recruited 203 all-comer patients with average stent diameter of 2.58 ± 0.22 mm with a device delivery success rate of 98.0 %. The patients were followed up for 6 months whereas the overall MACE rate was 13.0 % with cumulative TLR of 6.0 % (Strehblow et al. 2007) which was higher than the TLR rate in this study. In this context a word of caution is indicated when comparing clinical event rates across studies. Particularly in the case of small caliber vessels, restenosis may be asymptomatic and therefore difficult to be detected unless an angiographic follow-up is routinely conducted.

Table 4 Clinical outcomes

PCI in very small vessels are often considered controversial because (1) very small vessels are responsible for the perfusion of a small coronary territory (2) they may not be worthwhile for stenting (3) they have a higher risk of dissection, perforation (4) or restenosis (5) and they are technically more challenging in terms of lesion crossability. The clinical relevance of very small vessel CAD should be interpreted within the clinical framework of the individual patient. The procedural success rate improved significantly with novel designs of newer generation BMS. In our study, we have found that the technical success in delivering the BMS into very small vessels was on the same high level with 97.9 % as compared to the patient group with vessels >2.5 mm (97.7 %). A higher percentage of very small vessel lesions received pre-dilatation, likely because of higher calcification burden and diffuse vessel disease. However, stent inflation pressures did not differ in both groups.

One of the known advantages of BMS is the duration of DAPT post PCI. The risk of stent thrombosis in DES is often related to non-compliance or premature cessation of DAPT. Hence, if compliance to DAPT becomes a concern, BMS could be a reasonable option. In this study, we have shown that BMS performed equally well when implanted in very small vessels as compared larger vessels. The implication would be shorter duration of DAPT in patients undergoing elective PCI with BMS. However, this duration may need to be prolonged in patients undergoing PCI in the setting of ACS.

Furthermore, in special populations like the Asian patients, with higher prevalence of diabetes mellitus (Ahmad et al. 2013), the elderly and females usually have diffuse and very small vessel disease. The distal parts of the artery when the artery are very small, tortuous and calcified there may be an issue with deliverability of the stent for some patients, thus new generation thin strut BMS can be considered.

Limitations

This study has several limitations. First of all, this is a single-group, non-randomized design, which may have some degree of selection bias. The indication for BMS implantations may differ on a country basis depending on the risk profile for individual patients. Decisions may also have been impacted by local device costs in some countries. Data analysis is hence descriptive in nature and inferior to a randomized trial as no direct comparison can be made versus a control group. Secondly, the results reported here may have been affected by the type of bias inherent in all registries, namely the selective inclusion of lower-risk patients, together with less exhaustive monitoring than that applied in randomized controlled trials, potentially contributing to an overall under-reporting of events. Moreover, mandatory clinical, in-hospital follow-ups would have been preferred over telephone interviews; however, the regulatory ramification of this protocol modification would have rendered this study to be unfeasible. The follow-up rate of 93.9 % in this registry might be able to compensate for some of these deficits. Lastly the sample size calculation was based on the 9-month TLR rate of the overall population. Consequently, the descriptive nature of this subgroup analysis without the claim of proper statistical power must be mentioned as well.

Conclusion

Ultra-thin strut cobalt chromium BMS can be used in coronary artery disease (CAD) of very small reference diameters with very a high procedural success rates. A difference in clinical event rates despite differences in cardiovascular risk factors and lesion characteristics was not observed. However, PCI with DES remains the standard of care unless cost issues have an impact on the decision making process.

Abbreviations

ARC:

Academic Research Consortium

ACC:

American College of Cardiology

ACS:

acute coronary syndrome

AHA:

American Heart Association

BMS:

bare metal stent

CABG:

coronary artery bypass graft surgery

CAD:

coronary artery disease

DAPT:

dual anti-platelet therapy

DES:

drug eluting stent

DCB:

drug coated balloon

ECG:

electrocardiogram

ISR:

in stent restenosis

MACE:

major adverse cardiac event

PCI:

percutaneous coronary intervention

POBA:

plain old balloon angioplasty

STEMI:

ST elevation myocardial infarction

TLR:

target lesion revascularization

TIMI:

thrombolysis in myocardial infarction

References

  • Agostoni P, Biondi-Zoccai GG, Gasparini GL et al (2005) Is bare-metal stenting superior to balloon angioplasty for small vessel coronary artery disease? Evidence from a meta-analysis of randomized trials. Eur Heart J 26:881–889

    Article  PubMed  Google Scholar 

  • Ahmad WAW, Mohd Ali R, Khanom M et al (2013) The Journey of Malaysian NCVD-PCI (National cardiovascular Disease Database-Percutaneous Coronary Intervention) registry: a summary of three years report. Int J Cardiol 165:161–164

    Article  PubMed  Google Scholar 

  • Biondi-Zoccai G, Sangiorgi GM, Antoniucci D et al (2007) Taxus in Real-life Usage Evaluation Study. Testing prospectively the effectiveness and safety of paclitaxel-eluting stents in over 1000 very high-risk patients: design, baseline characteristics, procedural data and in-hospital outcomes of the multicenter Taxus in Real-life Usage Evaluation (TRUE) Study. Int J Cardiol 117:349–354

    Article  PubMed  Google Scholar 

  • Biondi-Zoccai G, Moretti C, Abbate A et al (2010) Percutaneous coronary intervention for small vessel coronary artery disease. Cardiovasc Revascularization Med 11(3):189–198

    Article  Google Scholar 

  • Bocksch W, Pomar F, Dziarmaga M et al (2010) Clinical safety and efficacy of a novel thin strut cobalt chromium coronary stent system: results of the real world Coroflex Blue registry. Catheter Cardiovasc Interv 75:78–85

    PubMed  Google Scholar 

  • Cutlip D, Windecker S, Mehran R (2007) Clinical end points in coronary stent trials: a case for standardized definitions. Circulation 115:2344–2351

    Article  PubMed  Google Scholar 

  • Elezi S, Kastrati A, Neumann FJ et al (1998) Vessel size and long-term outcome after coronary stent placement. Circulation 98:1875–1880

    Article  CAS  PubMed  Google Scholar 

  • Farooq V, Gogas BD, Serruys PW (2011) Restenosis: delineating the numerous causes of drug-eluting stent restenosis. Circ Cardio Inter 4:195–205

  • Kastrati A, Schomig A, Dirschinger J et al (2000) A randomized trial comparing stenting with balloon angioplasty in small vessels in patients with symptomatic coronary artery disease. Circulation 102:2593–2598

    Article  CAS  PubMed  Google Scholar 

  • Kastrati A, Mehili J, Dirschinger J et al (2001) Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO). Circulation 103:2816–2821

    Article  CAS  PubMed  Google Scholar 

  • Kirtane AJ, Gupta A, Iyengar S et al (2009) Safety and efficacy of drug-eluting and bare metal stents: comprehensive meta-analysis of randomized trials and observational studies. Circulation 119:3198–3206

    Article  CAS  PubMed  Google Scholar 

  • Koning R, Eltchaninoff H, Commeau P et al (2001) Stent placement compared with balloon angioplasty for small coronary arteries: in-hospital and 6-month clinical and angiographic results. Circulation 104:1604–1608

    Article  CAS  PubMed  Google Scholar 

  • Morice MC, Serruys PW, Sousa JE et al (2002) A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med 346:1773–1780

    Article  CAS  PubMed  Google Scholar 

  • Morton A, Papadopoulos T, Wales C et al (2003) ‘Real world’ small vessel coronary artery stenting: an analysis. Br J Cardiol (Acute Interv Cardiol) 10(1): AIC 28–AIC 32

  • Moses JW, Leon MB, Popma JJ et al (2003) Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med 349:1315–1323

    Article  CAS  PubMed  Google Scholar 

  • O’Connor NJ, Morton JR, Birkmeyer JD et al (1996) Effect of coronary artery diameter in patients undergoing coronary bypass surgery. Circulation 93:652–655

    Article  PubMed  Google Scholar 

  • Pache J, Kastrati A, Mehili J et al (2003) Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO-2). J Am Coll Cardiol 41:1283–1288

    Article  PubMed  Google Scholar 

  • Pache J, Dibra A, Mehili J et al (2006) Drug-eluting stents compared with thin-strut bare stents for the reduction of restenosis: a prospective, randomized trial. Eur Heart J 26:1262–1268

    Article  Google Scholar 

  • Park SW, Lee CW, Hong MK et al (2000) Randomized comparison of coronary stenting with optimal balloon angioplasty for treatment of lesions in small coronary arteries. Eur Heart J 21:1785–1789

    Article  CAS  PubMed  Google Scholar 

  • Park SJ, Shim WH, Ho DS et al (2003) A paclitaxel-eluting stent for the prevention of coronary restenosis. N Engl J Med 348:1537–1545

    Article  CAS  PubMed  Google Scholar 

  • Pfisterer M, Brunner-La Rocca HP, Buser PT et al (2006) Late clinical events after clopidogrel discontinuation may limit the benefit of drug-eluting stent: an observational study of drug-eluting stent restenosis. J Am Coll Cardiol 48:2584–2591

    Article  CAS  PubMed  Google Scholar 

  • Ryan TJ, Faxon DP, Funnar PM et al (1998) Guidelines for percutaneous transluminal coronary angioplasty: a report of the American College of Cardiology/American Heart Association task force on assessment of diagnostic and therapeutic cardiovascular procedures (subcommittee on percutaneous transluminal coronary angioplasty). J Am Coll Cardiol 12:529–545

    Article  Google Scholar 

  • Strehblow C, Gyongyosi M, Zenker G et al (2007) Small vessel stenting with cobalt-chromium stents (Arthos Pico) in a real-world setting. Coron Artery Dis 18:305–311

    Article  PubMed  Google Scholar 

  • Suttorp MJ, Stella PR, Dens J et al (2015) Ultra-thin strut cobalt chromium bare metal stent usage in a complex real-world setting: SOLSTICE registry. Neth Heart J 23:124–129

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Togni M, Eber S, Widmer J et al (2007) Impact of vessel size on outcome after implantation of sirolimus-eluting and paclitaxel-eluting stents: a subgroup analysis of the SIRTAX trial. JACC 50(12):1123–1131

    Article  CAS  PubMed  Google Scholar 

  • Windecker S, Kolh P, Alfonso F et al (2014) ESC/EACTS Guidelines on myocardial revascularization: the Task Force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS)Developed with the special contribution of the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur Heart J 35:254–619

    Google Scholar 

  • Wöhrle J, Zadura M, Möbius-Winkler S, Leschke M, Opitz C et al (2012) SeQuent please world wide registry: clinical results of sequent please paclitaxel-coated balloon angioplasty in a large-scale, prospective registry study. J Am Coll Cardiol 60:1733–1738

    Article  PubMed  Google Scholar 

  • Zeymer U, Waliszewski M, Spiecker M et al (2014) Prospective ‘real-world’ registry for the use of the ‘PCB only’ strategy in small vessel de novo lesions. Heart 100(4):311–316

    Article  CAS  PubMed  Google Scholar 

Download references

Authors’ contributions

MDI and WAWA are responsible for the main manuscript write-up and submission for publication. They are also actively involved in the data collection in their institution. ML, MW and MB are imperative in the overall study design and statistical analysis. However, they were not involved in the treatment of patients nor data collection. IZA and ASMZ contributed in the data collection and patient’s follow-up. They also provided intellectual contribution towards the manuscript write-up. All authors read and approved the final manuscript.

Acknowledgements

The authors would like to express their gratitude to Denny Herberger (Medical Scientific Affairs Berlin) and Ms. Aude Michaud (Medical Scientific Affairs France) for their logistic support to conduct this registry.

Competing interests

MW and MB are employees of Medical Scientific Affairs at B.Braun Vascular System. They have contributed in the overall design of this registry and data analysis. However, they had no role in the clinical decision making and the treatment of the subjects who participated in this registry. The other authors claimed no competing interest regarding this clinical research.

Availability of data and materials

The data are presented in tables provided together with the main manuscript. Should the journal require further details of the collected data; it can be obtained from the B.Braun Vascular System.

Consent for publication

Not applicable.

Ethics approval and consent to participate

In France this study was approved by the “Comité Consultative sur le Traitement de I’Information en matière de Recherche dans le domaine de la Santé (CTIRS dossier no. 12.384) and the Commission Nationale de I’informatique et des Libertés (CNIL, demande d’autorisation no. 912431). In Malaysia, this study received approval from University of Malaya Medical Centre ethics committee, Universiti Kebangsaan Malaysia ethics committee as well as from National Medical Research and Ethics Committee. The legal requirements are different in the participating countries. In as much this clinical assessment was an observational study under routine use, the conduct in Germany was regulated by the German Device law (Medizin Produkte Gesetz §23b) which did not require a formal ethics vote. Likewise in the Netherlands, Switzerland and Croatia, ethics votes were not required given that data collection was done in a pseudonymized manner (no patient initials, only birth years and not birth dates). Also these patients would have been treated with the device independent of the data collection. The clinical follow-up was part of the routine patient care established in each participating country. With the exception of France, two votes were mandatory for the collection of data and one for the ethical conduct of the study (please see the attached documents). The responsibility for the documentation of the local requirements was at the individual centers. The Declaration of Helsinki was respected in all countries. An informed consent text specific to the registry approved by the local ethics committee will have to be signed by the patients voluntarily prior to the enrolment.

Funding

There was no financial support provided by the above mentioned company to any of the institutions involved with regards to the treatment of the study patients.

Impact on daily practice

In ‘real-world’ practice, ultra-thin struts high end BMS with excellent crossing profiles may be considered as an alternative to DES in CAD with very small reference vessel diameters when deliverability, DAPT interruption, bleeding risks, or cost may be a concern.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Muhammad Dzafir Ismail.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ismail, M.D., Ahmad, W.A.W., Leschke, M. et al. The outcomes of patients with very small coronary artery disease treated with thin strut cobalt chromium bare metal stents: an observational study. SpringerPlus 5, 1668 (2016). https://doi.org/10.1186/s40064-016-3350-7