From: Cup implantation accuracy using the HipCOMPASS mechanical intraoperative support device

Fig. 4

a After the device is adjusted for the pelvic width, pelvic height, and error caused by soft tissue thickness, the legs of HipCOMPASS are placed on the left and right ASIS and on the pubic symphysis. In this position, the device is fixed by insertion of 1.5-mm K-wires through the hollow left and right ASIS legs and into the ASIS. An assistant applies appropriate force to the anterior of HipCOMPASS to hold the device in place, thereby allowing the indicator to show accurate angles. b An additional device is attached to the side of the cup holder with a bar (asterisk) parallel to the orientation of the cup (asterisks), and the cup is inserted such that it sits parallel to the device indicator. The bar and the indicator are set very close to each other, allowing the cup to be inserted more accurately