Skip to main content

Table 1 Back calculated concentrations of calibration standards

From: Development and validation of an HPLC method for the determination of vancomycin in human plasma and its comparison with an immunoassay (PETINIA)

Back calculation (n = 5)

Nominal concentration (mg/L)

60

30

10

1

0.5

0.25

Mean (mg/L)

58.0

30.2

10.3

0.99

0.49

0.25

SD

2.09

1.24

0.31

0.07

0.02

0.01

Recovery (%)

96.6

100.7

103.5

99.2

99.6

100.3

CV (%)

3.61

4.11

3.03

6.93

4.61

3.23