Table 5 Linear regression summary statistics of specimen comparisonsa

From: Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer

 

Slope (95 % CI)b

Y-intercept (95 % CI)b (mg/L)

Correlation coefficient

95 % CI at lower MDLc (mg/L)

95 % CI at upper MDLc (mg/L)

EDTA vs. serum

 Kappa

0.969 (0.950–0.988)

2.58 (1.54–3.64)

0.9954

4.8–6.8

20.4–22.4

 Lambda

0.995 (0.986–1.003)

0.367 (0.127–0.608)

0.9992

5.8–6.3

26.3–26.7

 κ/λ ratio

1.001 (0.982–1.021)

0.133 (0.066–0.199)

0.9957

0.33–0.46

1.73–1.84

Li-heparin plasma separator tube vs. serum

 Kappa

1.081 (1.063–1.100)

0.64 (−0.38 to 1.67)

0.9964

3.2–5.2

19.4–22.5

 Lambda

0.988 (0.981–1.000)

0.279 (−0.101 to 0.557)

1.0000

5.7–6.3

26.1–26.6

 κ/λ ratio

1.175 (1.154–1.196)

−0.091 (−0.188 to 0.007)

0.9961

0.12–0.31

1.65–1.83

  1. aAnalysis uses Deming regression excluding outliers (see Table 4)
  2. b CI confidence interval
  3. c MDL medical decision limit: kappa, 3.3 and 19.4 mg/L; lambda, 5.7 and 26.3 mg/L; κ/λ ratio, 0.26 and 1.65