Skip to main content

Table 2 Accuracy vancomycin ELISA

From: ELISA-based detection of gentamicin and vancomycin in protein-containing samples

Vancomycin concentration in validation series (ng/ml)

Concentration determined by polynomial regression (ng/ml)

Recovery (%)

31.3

33.0

105.6

62.5

68.0

108.8

125.0

114.5

91.6

250.0

258.6

103.4

750.0

766.3

102.2

4000.0

4282.1

107.1

Hum. wound exudate (50 ng/ml)

53.5

107.0