Skip to main content

Table 5 Sample determination results

From: Evaluation of the Di(2-ethylhexyl)phthalate released from polyvinyl chloride medical devices that contact blood

Medical devices

Extraction media

DEHP released, mg

Transfusion set for single usea

Water/ethanol mixture

2.84

blood

0.068-0.073b

Leukcyte-reducing filter for single use

Water/ethanol mixture

1.2

Tube system for haemodialysis use

Water/ethanol mixture

11.6

  1. a100 mL of the extraction media (both the water/ethanol mixture and blood) was used.
  2. bBlood from three healthy volunteers