Skip to main content

Table 2 Precision and accuracy at four different concentrations

From: Bioequivalance and pharmacokinetic study of febuxostat in human plasma by using LC-MS/MS with liquid liquid extraction method

Spiked plasma concen-tration (ng/mL)

Within-run

Between-run

Concentration measured

CVa(%)

Accuracy %

Concentration measured

CV.a(%)

Accuracy %

(n = 6)

(n = 30)

(ng/mL)

(ng/mL)

(mean ± S.d.)

(mean ± S.D.)

1.00

0.98 ± 0.04

3.88

98.43

1.03 ± 0.09

8.44

103.19

3.00

2.92 ± 0.11

3.68

97.33

3.02 ± 0.17

5.57

100.78

4000.00

3957.78 ± 104.45

2.64

98.94

4021.78 ± 117.78

2.93

100.54

4800.00

4754.29 ± 158.36

3.33

99.05

4814.48 ± 133.07

2.76

100.30

  1. a(Standard deviation/mean concentration measured)x100.