Table 2 Results obtained from the three scans performed with the PDM Si-SPAD detector at different z-positions

From: Improvement of the detection efficiency calibration and homogeneity measurement of Si-SPAD detectors

Scan number

z-position (mm)

x-center (mm)

y-center (mm)

Diameter (mm)

1

13.6

228.81

6.52

0.23

2

15.6

228.79

6.56

0.23

3

14.6

228.80

6.55

0.04

  1. The values (x, y and z) obtained from scan Nr. 3 are the final alignment positions where the Si-SPAD must be placed for its calibration. The diameter at this position corresponds approximately to the diameter (ϕ D = 50 μm) of the detector active area